High sensitivity CDC EUA SARS-CoV-2 kit-based End Point-PCR assay.

This article has 1 evaluations Published on
Read the full article Related papers
This article on Sciety

Abstract

SARS-CoV-2 diagnosis is based on RT-qPCR protocols that limited testing to facilities with Real Time PCR devices and probes supply. Here we described and adapted version of the RT-qPCR CDC protocol where N1, N2 and N3 primers are used for end point PCR and amplicons are visualized on agarose gel with a limit of detection up to 20 viral RNA copies/uL. This protocol would allow to extend SARS-CoV-2 diagnosis to basic molecular biology laboratories with a great impact on surveillance programs at developing countries.