Comparison of the Accula SARS-CoV-2 Test with a Laboratory-Developed Assay for Detection of SARS-CoV-2 RNA in Clinical Nasopharyngeal Specimens
Abstract
Background
Several point-of-care (POC) molecular tests have received emergency use authorization (EUA) from the Food and Drug Administration (FDA) for diagnosis of SARS-CoV-2. The test performance characteristics of the Accula (Mesa Biotech) SARS-CoV-2 POC test need to be evaluated to inform its optimal use.
Objectives
The aim of this study was to assess test performance of the Accula SARS-CoV-2 test.
Study design
The performance of the Accula test was assessed by comparing results of 100 nasopharyngeal swab samples previously characterized by the Stanford Health Care EUA laboratory-developed test (SHC-LDT) targeting the envelope ( E ) gene. Assay concordance was assessed by overall percent agreement, positive percent agreement (PPA), negative percent agreement (NPA), and Cohen’s kappa coefficient.
Results
Overall percent agreement between the assays was 84.0% (95% confidence interval [CI] 75.3 to 90.6%), PPA was 68.0% (95% CI 53.3 to 80.5%) and the kappa coefficient was 0.68 (95% CI 0.54 to 0.82). Sixteen specimens detected by the SHC-LDT were not detected by the Accula test, and showed low viral load burden with a median cycle threshold value of 37.7. NPA was 100% (95% CI 94.2 to 100%).
Conclusion
Compared to the SHC-LDT, the Accula SARS-CoV-2 test showed excellent negative agreement. However, positive agreement was low for samples with low viral load. The false negative rate of the Accula POC test calls for a more thorough evaluation of POC test performance characteristics in clinical settings, and for confirmatory testing in individuals with moderate to high pre-test probability of SARS-CoV-2 who test negative on Accula.