Evaluation of two commercial and two non-commercial immunoassays for the detection of prior infection to SARS-CoV-2

This article has 1 evaluations Published on
Read the full article Related papers
This article on Sciety

Abstract

Background

Seroepidemiology is an important tool to characterize the epidemiology and immunobiology of SARS-CoV-2 but many immunoassays have not been externally validated raising questions about reliability of study findings. To ensure meaningful data, particularly in a low seroprevalence population, assays need to be rigorously characterized with high specificity.

Methods

We evaluated two commercial (Roche Diagnostics and Epitope Diagnostics IgM/IgG) and two non-commercial (Simoa and Ragon/MGH IgG) immunoassays against 68 confirmed positive and 232 pre-pandemic negative controls. Sensitivity was stratified by time from symptom onset. The Simoa multiplex assay applied three pre-defined algorithm models to determine sample result.

Results

The Roche and Ragon/MGH IgG assays each registered 1/232 false positive, the primary Simoa model registered 2/232 false positives, and the Epitope registered 2/230 and 3/230 false positives for the IgG and IgM assays respectively. Sensitivity >21 days post symptom-onset was 100% for all assays except Epitope IgM, but lower and/or with greater variability between assays for samples collected 9-14 days (67-100%) and 15-21 days (69-100%) post-symptom onset. The Simoa and Epitope IgG assays demonstrated excellent sensitivity earlier in the disease course. The Roche and Ragon/MGH assays were less sensitive during early disease, particularly among immunosuppressed individuals.

Conclusions

The Epitope IgG demonstrated good sensitivity and specificity. The Roche and Ragon/MGH IgG assays registered rare false positives with lower early sensitivity. The Simoa assay primary model had excellent sensitivity and few false positives.

Summary

SARS-CoV-2 immunoassays can be valuable tools for informing the global response, but many currently available assays have not been independently validated. We conducted a performance assessment of four assays including the Roche Diagnostics and Epitope Diagnostics immunoassays.