Retrospective Clinical Evaluation of Four Lateral Flow Assays for the Detection of SARS-CoV-2 Antibodies

This article has 1 evaluations Published on
Read the full article Related papers
This article on Sciety

Abstract

Coronavirus disease 2019 (COVID-19) is a potentially life-threatening respiratory infection caused by severe acute respiratory coronavirus 2 (SARS-CoV-2), for which numerous serologic assays are available. In a CLIA laboratory setting, we used a retrospective sample set (n = 457) to evaluate two lateral flow immunoassays (LFIAs; two iterations of Rapid Response™ COVID-19 Test Cassette, BTNX Inc.) and a subset of to evaluate SARS-COV-2 IgG/IgM Rapid Test, ACON Laboratories (n = 200); and Standard Q COVID-19 IgM/IgG Duo, SD BIOSENSOR (n = 155) for their capacity to detect of SARS-CoV-2 IgG. In a cohort of primarily hospitalized patients with RT-PCR confirmed COVID-19, the BTNX assays demonstrated 95% and 92% agreement with the Abbott SARS-CoV-2 IgG assay and sensitivity was highest at ≥ 14 days from symptom onset [BTNX kit 1, 95%; BTNX kit 2, 91%]. ACON and SD assays demonstrated 99% and 100% agreement with the Abbott assay at ≥ 14 days from symptom onset. Specificity was measured using 74 specimens collected prior to SARS-CoV-2 circulation in the United States and 31 “cross-reactivity challenge” specimens, including those from patients with a history of seasonal coronavirus infection and was 98% for BTNX kit 1 and ACON and 100% for BTNX kit 2 and SD. Taken with data from EUA assays, these results suggest that LFIAs may provide adequate results for rapid detection of SARS-CoV-2. Replicating these results in fingerstick blood in outpatient populations, would further support the possibility that LFIAs may be useful to increase access to serologic testing

Related articles

Related articles are currently not available for this article.