Comparison of RNA extraction methods for the detection of SARS-CoV-2 by RT-PCR
Abstract
Objectives
The SARS-CoV-2 pandemic outbreak has stressed health care systems as well as medical supply chains, but diagnostic testing is an essential public health measure to control viral spread. Here we test the suitability of different RNA extraction methods for integration into a diagnostic workflow for coronavirus testing.
Methods
We applied six RNA extraction methods on the same 24 SARS-CoV-2 positive patient samples and quantified their results by subsequent reverse-transcriptase PCR (RT-PCR) of three viral genes. These methods included a) column-based extraction, b) phenol-chloroform extraction, as well as c) extraction using magnetic beads (i.e., one commercial kit as well as three different magnetic beads in combination with home-brewed buffers and solutions).
Results
We achieved diagnostic-quality RT-PCR results with all methods, and there was no significant difference between the tested methods, except for one magnetic bead protocol with home-brewed buffers, in which the number of positive tested genes was significantly lower.
Conclusions
Five of the six RNA extraction methods are interchangeable in a diagnostic workflow. Since some methods are more scalable than others, and have comparable results on RT-PCR quantitation, they may be more amenable to high-throughput sample processing pipelines.
Graphical abstract
<fig id="ufig1" position="float" fig-type="figure" orientation="portrait"> <graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="20172494v1_ufig1" position="float" orientation="portrait"/> </fig>Related articles
Related articles are currently not available for this article.