Clinical validation of a point-of-care antibody test for COVID-19

This article has 1 evaluations Published on
Read the full article Related papers
This article on Sciety

Abstract

The objective of this study was to evaluate the performance of a lateral flow antibody test for COVID-19, approved for use in India. Although many point-of-care antibody tests are available globally, they have been subjected to limited clinical validation. This has led to suboptimal outcomes in the field, where antibody tests play a significant role in tracking the immunity of individuals and communities. In this study an antibody test, ImmunoQuick that recognizes antibodies to the Nucleocapsid and Spike proteins of SARS CoV-2 was tested in 100 symptomatic patients with a positive or negative diagnosis of COVID-19, based on RT-PCR results. The overall sensitivity of the test was found to be 86.1% (95% CI: 76.4% to 92.8%) and specificity 100% (95% confidence interval: 73.5% to 100%). The sensitivity reached a peak of 95.7% with samples taken 17 days after the onset of symptoms. Overall, the sensitivity and specificity of the test are sufficient for assessing seroprevalence.