Diagnostic Performance and Characteristics of Anterior Nasal Collection for the SARS-CoV-2 Antigen Test: A Prospective Study in Japan

This article has 1 evaluations Published on
Read the full article Related papers
This article on Sciety

Abstract

We conducted a prospective study in Japan to evaluate the diagnostic performance of the antigen test QuickNavi-COVID19 Ag using anterior nasal samples and to compare the degrees of coughs or sneezes induction and the severity of pain between anterior nasal collection and nasopharyngeal collection. A total of 862 participants were included in the analysis. In comparison to the findings of reverse transcription PCR using nasopharyngeal samples, the antigen test using anterior nasal samples showed 72.5% sensitivity (95% confidence interval [CI]: 58.3%-84.1%) and 100% specificity (95% CI: 99.3%-100%). Anterior nasal collection was associated with a significantly lower degree of coughs or sneezes induction and the severity of pain in comparison to nasopharyngeal collection ( p < 0.001). The antigen test using anterior nasal samples showed moderate sensitivity but was less painful and induced fewer coughs or sneezes.