Detection of SARS-CoV-2 variants by Abbott molecular, antigen, and serological tests
Abstract
Background
Viral diversity presents an ongoing challenge for diagnostic tests, which need to accurately detect all circulating variants. The Abbott Global Surveillance program monitors severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) variants and their impact on diagnostic test performance.
Objectives
To evaluate the capacity of Abbott molecular, antigen, and serologic assays to detect the SARS-CoV-2 B.1.1.7, B.1.351 and the P.1 variants.
Study design
Virus variant culture stock dilutions (B.1.1.7, BEI NR-54011; B.1.351, BEI NR-54008 and 54009; P.1, BEI NR-54982) and clinical samples from patients with confirmed B.1.1.7 variant infection were run on the Abbott ID NOW COVID-19, m 2000 RealTi m e SARS-CoV-2, Alinity m SARS-CoV-2, and Alinity m Resp-4-Plex molecular assays; the BinaxNOW COVID-19 Ag Card and Panbio COVID-19 Ag Rapid Test Device; and the ARCHITECT/Alinity i SARS-CoV-2 IgG and AdviseDx IgM assays, Panbio COVID-19 IgG assay, and ARCHITECT/Alinity i AdviseDx SARS-CoV-2 IgG II assay.
Results
Cultured virus stocks and B.1.1.7 clinical samples were detected with molecular, antigen, and serologic assays in the expected ranges, confirming in silico predictions. The ratio between genome equivalents (GE) and calculated median tissue culture infectious dose (TCID 50 ) were 31-to 83-fold higher for B.1.1.7 cultures compared to B.1.351 and P.1 cultures, demonstrating that GE are more consistent units between cultures than TCID 50 .
Conclusions
Abbott molecular and antigen assays effectively detect B.1.1.7, B.1.351, and P.1 variant infections and Abbott serologic assays detect B.1.1.7 antibodies in patient sera. Future studies with SARS-CoV-2 virus cultures should use quantitative viral load values to compare detection of variants.
Highlights
-
Abbott SARS-CoV-2 molecular and antigen assays detect B.1.1.7, B.1.351, and P.1 variants
-
Abbott SARS-CoV-2 antibody assays detect B.1.1.7 antibodies in recovered patient sera
-
Quantitation of viral load in genome equivalents allows comparison of assay performance