Diagnostic performance of a novel digital immunoassay (RapidTesta SARS-CoV-2): a prospective observational study with 1,127 nasopharyngeal samples
Abstract
Introduction
Digital immunoassays are generally regarded as superior tests for the detection of infectious disease pathogens, but there have been insufficient data concerning SARS-CoV-2 immunoassays.
Methods
We prospectively evaluated a novel digital immunoassay (RapidTesta SARS-CoV-2). Two nasopharyngeal samples were simultaneously collected for antigen tests and RT-PCR. Real-time RT-PCR for SARS-CoV-2, using a method developed by the National Institute of Infectious Diseases, Japan, served as the reference RT-PCR method.
Results
During the study period, 1,127 nasopharyngeal samples (symptomatic patients: 802, asymptomatic patients: 325) were evaluated. For digital immunoassay antigen tests, the sensitivity was 78.3% (95% CI: 67.3%–87.1%) and the specificity was 97.6% (95% CI: 96.5%–98.5%). When technicians visually analyzed the antigen test results, the sensitivity was 71.6% (95% CI: 59.9%–81.5%) and the specificity was 99.2% (95% CI: 98.5%–99.7%). Among symptomatic patients, the sensitivity was 89.4% (95% CI; 76.9%–96.5%) with digital immunoassay antigen tests, and 85.1% (95% CI; 71.7%–93.8%) with visually analyzed the antigen test, respectively.
Conclusions
The findings indicated that RapidTesta SARS-CoV-2 analysis with the DIA device had sufficient analytical performance for the detection of SARS-CoV-2 in nasopharyngeal samples. When positive DIA results are recorded without a visually recognizable red line at the positive line location on the test cassette, additional RT-PCR evaluation should be performed.
Related articles
Related articles are currently not available for this article.